The need to update NCRP 151 data for 10 MV linear accelerator bunker shielding based on new measurements and Monte Carlo simulations.
James RijkenS TownsB HealyPublished in: Journal of radiological protection : official journal of the Society for Radiological Protection (2021)
Linear accelerator bunker shielding protocols such as NCRP 151 have previously been tested against a large sample of measured data from real bunkers and machines but differences in per-energy concrete penetration (TVLs) for 10 MV has not yet been resolved. These differences are likely due to historical beam data and can potentially result in over-exposure of radiation workers and the public. This study examines a cohort of clinical linac bunker survey measurements and compares them to popular shielding protocols. Differences were investigated using contemporary beam data for both Monte Carlo simulation and in analytical equations. For primary barriers, NCRP 151 underestimates the dose rate through concrete by on average a factor of 2 with secondary barriers and maze entrance doses having much better agreement. Use of contemporary beam data in Monte Carlo simulation and an analytical equation yielded TVL values much closer to the measured values compared to NCRP 151. The TVL data in NCRP 151 is outdated and needs to be updated based upon the energy spectra of modern linear accelerators. Until then, physicists should use the TVL values shown in this study instead.