How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment.
Martin ThomasRiddhi JoshiGrant CavePublished in: Critical care research and practice (2021)
A randomised benchtop experiment was conducted, with a size 8 uncuffed Portex tracheostomy connected to an Optiflow™ with Airvo 2™ humidifier system. The tracheostomy tube was partially immersed in water to give rise to a column of water within the inner surface of the tube. An air fluid interface was generated with flows of 40 L/min, 50 L/min, and 60 L/min. The amount of potential PEEP (pPEEP) generated was determined by the distance the water column was pushed downward by the flow delivered. Findings. Overall 40 L/min, 50 L/min, and 60 L/min provided pPEEP of approximately 0.3 cmH2O, 0.5 cmH2O, and 0.9 cmH2O, respectively. There was a statistically significant change in pPEEP with change in flows from 40-60 L/min with an average change in pPEEP of 0.25-0.35 cmH2O per 10 L/min flow (p value <0.01). Interpretation. HFT can generate measurable and variable PEEP despite the open system used. The pPEEP generated with HFT is minimal despite statistically significant change with increasing flows. This pPEEP is unlikely to provide mechanical benefit in weaning patients off ventilatory support.