Healing Abutment Distortion in Implant Prostheses: An In Vitro Study.
Chandrashekhar PandeyBishwa Prakash BhattaraiApiwat RiddhabhayaNatthamet WongsirichatDinesh RokayaPublished in: Journal of functional biomaterials (2022)
Little has been studied regarding the repeated usage of healing abutments and their effects on the distortion of dental implant-healing abutment hex surfaces. Additionally, implant manufacturers do not provide specific guidelines on how many times a healing abutment can be used before discarding. Therefore, we evaluated the effect of repeated screwing-unscrewing of implant-healing abutment on the healing abutment hex surface and screw head. A total of 12 Biomate implants with 4 mm diameter and 13 mm length were inserted into a synthetic bone block. The standard healing abutments of 3 mm diameter and 4 mm length were screwed onto each implant using a torque ratchet at a final torque of 30 Ncm. Immediately, the abutments were unscrewed at 30 Ncm. Then, screwing-unscrewing was repeated for 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 times and the healing abutments were scanned under the scanning electron microscope for any distortion. Distortion was graded as 0, 1, 2, and 3. Data were analyzed using SPSS 24.0. Descriptive statistics were calculated. One-way ANOVA with post hoc using Tukey's HSD test was performed to analyze the difference in distortion at different screwing-unscrewing times. A significant level was selected at p -value = 0.05. It was found that distortion healing abutments screws were seen after 32 times screwing and unscrewing. There was a significant difference in the distortion ( p -value < 0.05) after 24 times of repeated usage of healing abutment and at 160, 320, and 400 times. No surface distortions were observed at the healing abutment screw head at 4, 8, 16, 24, 32, 40, 80, 160, 320, and 400 cycles of screwing-unscrewing. It can be concluded that repeated screwing and unscrewing of the implant-healing abutments causes damage to the healing abutment hex surface. The distortion of healing abutments screws was seen after 32 times screwing and unscrewing. No surface distortions were observed on the healing abutment screw head until 400 times of screwing and unscrewing. Hence, the clinician should be cautious while using the healing abutments repeatedly.