Radiation exposure during cardiac device implantation: Lessons learned from a multicenter registry.

Paul BruAntoine DompnierWalid AmaraGeorges HaddadGianina GaluscanPascal SagnolMathieu SteinbachChristian MontagnierJérôme TaiebJulien BeguecAntoine Milhemnull null
Published in: Pacing and clinical electrophysiology : PACE (2019)
Based on real-life measurements, this multicenter registry provides dose reference levels and may help centers assess radiation exposure. Although biventricular device implantation was responsible for the highest radiation exposure, FT was meaningfully shortened compared to previously reported values. For a same FT, the use of new generators and custom settings has significantly reduced DAP.
Keyphrases
  • cross sectional
  • double blind
  • left ventricular