Calibration-free pTx of the human heart at 7T via 3D universal pulses.
Christoph Stefan AignerSebastian DietrichTobias SchaeffterSebastian SchmitterPublished in: Magnetic resonance in medicine (2021)
UP22-4kT allows for calibration-free homogeneous flip angle distributions across the human heart at 7T. Large inter-subject variations because of sex, age, and body mass index are well tolerated. The proposed universal pulse removes the need for lengthy (10-15 min) calibration scans and therefore has the potential to bring body imaging at 7T closer to the clinical application.