Automated closed volume reduction process for apheresis stem cell grafts: From development to clinical implementation.
Anita HowellBrenda LetcherKelly MurphyHeidi ElmoazzenTanya PetraszkoJason P AckerNicolas PineaultJelena L HolovatiPublished in: Transfusion (2022)
The effectiveness outcomes of the first Canadian experience in the implementation of a closed semi-automated volume reduction system in the processing of HPC-A products for autologous transplant have met the predetermined acceptance criteria, supporting its use in a stem cell manufacturing laboratory compliant with good manufacturing practice regulations.