Login / Signup

Continuous Renal Replacement Therapy During Long-term Normothermic Machine Perfusion of Human Donor Livers for up to 7 D.

Bianca LascarisRoland F HoffmannMaarten W N NijstenRobert J Porte
Published in: Transplantation direct (2024)
The integration of CRRT into the NMP system corrected the composition of the perfusate to near-physiological values, which could be maintained for up to 7 d. The ultrafiltrate can serve as an alternative to the perfusate to monitor concentrations of small molecules without potentially compromising sterility.
Keyphrases
  • endothelial cells
  • acute kidney injury
  • induced pluripotent stem cells
  • deep learning
  • pluripotent stem cells
  • magnetic resonance
  • computed tomography
  • neural network