Validation in the General Population of the iHealth Track Blood Pressure Monitor for Self-Measurement According to the European Society of Hypertension International Protocol Revision 2010: Descriptive Investigation.

Victoria Mazoteras PardoRicardo Becerro de Bengoa-VallejoMarta Elena Losa IglesiasDaniel López-LópezPatricia Palomo-LópezDavid Rodriguez SanzCésar Calvo Lobo
Published in: JMIR mHealth and uHealth (2019)
iHealth Track readings differed from the standard by less than 5, 10, and 15 mm Hg, fulfilling ESH-IP2 requirements. Consequently, this device is suitable for use in the general population.