Systemic IgG exposure and safety in patients with primary immunodeficiency: a randomized crossover study comparing a novel investigational wearable infusor versus the Crono pump.
Richard L WassermanCharlotte Cunningham-RundlesJohn AndersonPatricia LugarMichael PalumboNiraj C PatelJutta HofmannNathan PacoureauEileen RogersMichaela PrausMikhail A RojavinPublished in: Immunotherapy (2022)
Aim: A novel, Investigational Wearable Infusor (IWI) was evaluated in a randomized, controlled, crossover, open-label study to determine if its delivery of subcutaneous immunoglobulin (IgPro20) achieved a comparable area under the concentration-time curve (AUC) for immunoglobulin G (IgG) versus the Crono S-PID-50 infusion pump (CP). EudraCT: 2016-003798-16. Materials & methods: Patients with primary immunodeficiency (PID) were randomized to receive IgPro20 in Sequence 1 (CP/IWI) or 2 (IWI/CP). The primary end point was AUC for IgG during the final week of each 4-week period. Results: 23 patients were enrolled. Evaluation of area under the concentration-time curve from time 0 (pre-infusion) to 7 days after infusion (AUC 0-7 days ) (IWI: 1806 h*g/l; CP: 1829 h*g/l) and geometric mean ratio indicated comparable AUCs for IgG for both devices. Conclusion: Similar IgG exposure, indicated by AUC values, can be achieved with IgPro20 using the IWI or CP in PID.