Login / Signup

Modeling and Control of an MR-Safe Pneumatic Radial Inflow Motor and Encoder (PRIME).

Anthony L GundermanMilad AzizkhaniSaikat T SenguptaKevin R ClearyYue Chen
Published in: IEEE/ASME transactions on mechatronics : a joint publication of the IEEE Industrial Electronics Society and the ASME Dynamic Systems and Control Division (2023)
Magnetic resonance (MR) conditional actuators and encoders are the key components for MR-guided robotic systems. In this article, we present the modeling and control of our MR-safe pneumatic radial inflow motor and encoder. A comprehensive model is developed that considers the primary dynamic elements of the system, including: 1) motor dynamics, 2) pneumatic transmission line dynamics, and 3) valve dynamics. After model validation, we present a simplified third order model that facilitates design of a first order sliding mode controller (TO-SMC). Finally, the motor hardware is tested in a 7T MRI. No image distortion or artifacts were observed. We posit the MR-safe motor and dynamic model will lower the entry barriers for researchers interested in MR-guided robots and promote wider adoption of MR-guided robotic systems.
Keyphrases
  • contrast enhanced
  • magnetic resonance
  • magnetic resonance imaging
  • computed tomography
  • minimally invasive
  • aortic valve
  • heart failure
  • deep learning
  • robot assisted
  • left ventricular
  • electronic health record
  • cone beam