Design of a fast field-cycling magnetic resonance imaging system, characterization and methods for relaxation dispersion measurements around 1.5 T.

Nicolas ChanetGeneviève GuillotGeorges WilloquetLaurène JourdainRose-Marie DubuissonGaël ReganhaLudovic de Rochefort
Published in: The Review of scientific instruments (2020)
The dependence of the nuclear magnetic resonance relaxation rate on the magnetic field has been widely studied, in particular, in biomedical areas with the objectives to better understand the underlying microscopic mechanisms in tissues and provide biomarkers of diseases. By combining fast-field cycling (FFC) and magnetic resonance imaging (MRI), it is possible to provide localized relaxation dispersion measurements in heterogeneous systems with recent demonstrations in solutions, biological samples, human beings, and small animals. We report here the developments and performances of a device designed for small animal FFC-MRI comprising a resistive insert technology operating inside a 1.5 T MRI system. Specific measurement methods were developed to characterize the system efficiency, response time, homogeneity, stability, and compensation. By adding a non-linear element in the system and using a dual amplifier strategy, it is shown that large field offsets can be produced during relaxation periods while maintaining precise field control during detection periods. The measurement of longitudinal nuclear magnetic relaxation dispersion (NMRD) profiles in the range of 1.08 T-1.92 T is reported, essentially displaying a linear variation in this range for common MRI contrast agents. The slopes of both the longitudinal and transverse relaxation dispersion profiles at 1.5 T are measured and validated, extending the capabilities of previous approaches. The performances of a longitudinal relaxation dispersion mapping method are finally reported, opening the way to quantitative preclinical dispersion imaging studies at a high FFC-MRI field.
Keyphrases