Login / Signup

Computer simulation help predict the frame deformation following a Venus-A transcatheter aortic valve implantation in patients with pure aortic regurgitation: a retrospective study.

Feicheng YuSongzan ChenLili TianXulin HongBei WangLili WuYan MaGuosheng Fu
Published in: Cardiovascular diagnosis and therapy (2024)
Patient-specific computer simulation of TAVI accurately predicted frame deformation in Chinese patients with pure AR. The forces at the STJ facilitated stabilization of the device within the aortic root, which might be used as a discriminator to identify patients at risk of device migration prior to intervention.
Keyphrases