Login / Signup

Robotic middle ear access for cochlear implantation: First in man.

Marco CaversaccioWilhelm WimmerJuan AnsóGeorgios MantokoudisNicolas GerberChristoph RathgebDaniel SchneiderJan HermannFranca WagnerOlivier ScheideggerMarkus HuthLukas AnschuetzMartin KompisTom WilliamsonBrett BellKate GavaghanStefan Weber
Published in: PloS one (2019)
To demonstrate the feasibility of robotic middle ear access in a clinical setting, nine adult patients with severe-to-profound hearing loss indicated for cochlear implantation were included in this clinical trial. A keyhole access tunnel to the tympanic cavity and targeting the round window was planned based on preoperatively acquired computed tomography image data and robotically drilled to the level of the facial recess. Intraoperative imaging was performed to confirm sufficient distance of the drilling trajectory to relevant anatomy. Robotic drilling continued toward the round window. The cochlear access was manually created by the surgeon. Electrode arrays were inserted through the keyhole tunnel under microscopic supervision via a tympanomeatal flap. All patients were successfully implanted with a cochlear implant. In 9 of 9 patients the robotic drilling was planned and performed to the level of the facial recess. In 3 patients, the procedure was reverted to a conventional approach for safety reasons. No change in facial nerve function compared to baseline measurements was observed. Robotic keyhole access for cochlear implantation is feasible. Further improvements to workflow complexity, duration of surgery, and usability including safety assessments are required to enable wider adoption of the procedure.
Keyphrases