On the basis of the knowledge gained from this study on the test-retest reliability of pulsed DPOAE signals and the current literature, we propose a DPOAE protocol for future serial monitoring applications that takes into account the following factors: (1) separation of DPOAE components, (2) use of individually optimal stimulus parameters, (3) SNR of at least 15 dB, (4) accurate pressure calibration, (5) consideration of frequency- and level-dependent test-retest reliabilities and corresponding reference ranges, and (6) stimulus levels L2 that are as low as possible with sufficient SNR to capture the nonlinear functional state of the cochlear amplifier operating at its highest gain.