Assessing mechanical catheter dysfunction in automated tidal peritoneal dialysis using cycler software: a case control, proof-of-concept study.
Krystell Oviedo FloresLukas KalteneggerFabian EibensteinerMarkus UnterwurzacherKlaus KratochwillChristoph AufrichtFranz KoenigAndreas VychytilPublished in: Scientific reports (2022)
New recommendations on evaluation of peritoneal membrane function suggest ruling out catheter dysfunction when evaluating patients with low ultrafiltration capacity. We introduce the use of a combination of parameters obtained from the cycler software PD Link with HomeChoicePro (Baxter International Inc., Illinois, United States) cyclers for predicting catheter dysfunction in automated peritoneal dialysis patients (APD). Out of 117 patients treated at the Medical University of Vienna between 2015 and 2021, we retrospectively identified all patients with verified catheter dysfunction (n = 14) and compared them to controls without clinical evidence of mechanical catheter problems and a recent X-ray confirming PD catheter tip in the rectovesical/rectouterine space (n = 19). All patients had a coiled single-cuff PD catheter, performed tidal PD, and received neutral pH bicarbonate/lactate-buffered PD fluids with low-glucose degradation products on APD. Icodextrin-containing PD fluids were used for daytime dwells. We retrieved cycler data for seven days each and tested parameters' predictive capability of catheter dysfunction. Total number of alarms/week > 7 as single predictive parameter of catheter dislocation identified 85.7% (sensitivity) of patients with dislocated catheter, whereas 31.6% (1-specificity) of control patients were false positive. A combination of parameters (number of alarms/week > 7, total drain time > 22 min, ultrafiltration of last fill < 150 mL) where at least two of three parameters appeared identified the same proportion of patients with catheter dislocation, but was more accurate in identifying controls (21.1% false positive). In contrast to yearly PET measurements, an easily applicable combination of daily cycler readout parameters, also available in new APD systems connected to remote monitoring platforms shows potential for diagnosis of catheter dysfunction during routine follow-up.
Keyphrases
- end stage renal disease
- peritoneal dialysis
- chronic kidney disease
- ultrasound guided
- ejection fraction
- oxidative stress
- newly diagnosed
- machine learning
- healthcare
- randomized controlled trial
- type diabetes
- computed tomography
- magnetic resonance
- magnetic resonance imaging
- blood pressure
- adipose tissue
- high throughput
- clinical trial
- metabolic syndrome
- depressive symptoms
- obstructive sleep apnea
- patient reported outcomes
- pet ct
- clinical practice
- weight loss
- sleep quality
- blood glucose