Comparison of Bruker Biotyper ® and Vitek ® MS matrix-assisted laser desorption/ionization-time-of-flight mass spectrometry platforms for the identification of filamentous fungi.
Julia D HankinsMegan H Amerson-BrownCameron August BrownAshleigh N RieglerKenneth L MuldrewJames J DunnPublished in: Future microbiology (2023)
Aims: To evaluate the performance of two matrix-assisted laser desorption/ionization-time-of-flight mass spectrometry platforms to identify molds isolated from clinical specimens. Methods: Fifty mold isolates were analyzed on Bruker Biotyper ® and Vitek ® MS platforms. Two Bruker Biotyper extraction protocols were assessed alongside the US FDA-approved extraction protocol for Vitek MS. Results: The Bruker Biotyper modified NIH-developed extraction protocol correctly identified more isolates than Bruker's protocol (56 vs 33%). For species in the manufacturers' databases, Vitek MS correctly identified 85% of isolates, with 8% misidentifications. The Bruker Biotyper identified 64%, with no misidentifications. For isolates not in the databases, the Bruker Biotyper did not misidentify any, and Vitek MS misidentified 36%. Conclusion: Both the Vitek MS and Bruker Biotyper accurately identified the fungal isolates, however Vitek MS was more likely to misidentify isolates than the Bruker Biotyper.