Comparison of the PneumoGenius® and RealStar® Pneumocystis jirovecii PCR CE-IVD assays with a lab developed test for the detection of Pneumocystis jirovecii.

Roel H T NijhuisPeggy C R GodschalkJorike H I SminkCindy van der ZeeErik J van Hannen
Published in: Medical mycology (2021)
Pneumocystis jirovecii (Pj) is a fungal pathogen that can cause severe and potential fatal pneumonia (Pneumocystis pneumonia, PCP) in immunocompromised patients. Microbiological diagnosis is necessary to confirm PCP, for which mainly real-time PCR assays are used by detecting Pj from bronchoalveolar lavage (BAL) specimens. In this study, we evaluate the performance of the CE-IVD PneumoGenius® assay and CE-IVD RealStar® Pneumocystis jirovecii PCR assay in comparison to the lab developed test (LDT) that is used in routine diagnostics. Comparison was done by including 100 BAL specimens: 25 retrospective specimens, selected based on results obtained with LDT (15 positive/10 negative), and 75 prospectively collected specimens. LDT (targeting MSG) was performed according to local procedures and the PneumoGenius® (targeting mtLSU and DHPS fas) and RealStar® assays (targeting mtLSU) according to manufacturer's instructions. Combining results of retrospective and prospective analysis, sensitivity was 69.7%, 100% and 100% for the LDT, PneumoGenius® and RealStar®, respectively. Specificity was 100% for LDT and Pneumogenius®, whereas RealStar® showed a specificity of 97%. Correlation of fungal loads found with the PneumoGenius® and RealStar® assays was high (R2: 0.98). The PneumoGenius® and RealStar® assays performed comparable, and both showed high sensitivity in comparison to the LDT. For optimal diagnosis of PCP, the LDT has to be replaced by another, more sensitive assay.
Keyphrases