Comparison of Low-Level to High-Level Disinfection in Eliminating Microorganisms From Ultrasound Transducers Used on Skin: A Noninferiority Randomized Controlled Trial.

Nathan PetersFrances WilliamsonMichelle J BauerStacey LlewellynPeter James SnellingNicole MarshPatrick N A HarrisAdam G StewartClaire M Rickard
Published in: Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine (2023)
Disinfection with LLD is noninferior to HLD when microorganisms from skin have contaminated the transducer. Therefore, using LLD for US transducers involved in percutaneous procedures would present no higher infection risk compared with HLD.