Proposed DRLs for mammography in Switzerland.
Laura DupontChristoph AberleDiomidis BotsikasMichael IthThiago Viana Miranda LimaRoman MenzPascal MonninPierre-Alexandre PolettiStefano PresillaAlexander SchegererLiana-Cristina StoicaPhilipp TruebMarta Sans MercePublished in: Journal of radiological protection : official journal of the Society for Radiological Protection (2024)
The aim of this study is to propose diagnostic reference levels (DRLs) values for mammography in Switzerland. For the data collection, a survey was conducted among a sufficient number of centres, including five University hospitals, several cantonal hospitals, and large private clinics, covering all linguistic regions of Switzerland to be representative of the clinical practice. The data gathered contained the mean glandular dose (MGD), the compressed breast thickness (CBT), the mammography model and the examination parameters for each acquisition. The data collected was sorted into the following categories: 2D or digital breast tomosynthesis (DBT) examination, craniocaudal (CC) or mediolateral oblique (MLO) projection, and eight categories of CBT ranging from 20 mm to 100 mm in 10 mm intervals. A total of 24 762 acquisitions were gathered in 31 centres on 36 mammography units from six manufacturers. The analysis showed that the data reflects the practice in Switzerland. The results revealed that the MGD is larger for DBT than for 2D acquisitions for the same CBT. From 20-30 mm to 90-100 mm of CBT, the 75th percentile of the MGD values obtained increased from 0.81 mGy to 2.55 mGy for 2D CC acquisitions, from 0.83 mGy to 2.96 mGy for 2D MLO acquisitions, from 1.22 mGy to 3.66 mGy for DBT CC acquisitions and from 1.33 mGy to 4.04 mGy for DBT MLO acquisitions. The results of the survey allow us to propose Swiss DRLs for mammography according to the examination type (2D/DBT), projection (CC/MLO) and CBT. The proposed values are very satisfactory in comparison with other studies.