Evaluation of image quality with four positron emitters and three preclinical PET/CT systems.

Jarmo TeuhoLeon RiehakainenAake HonkaniemiOlli MoisioChunlei HanMarko TirriShihao LiuTove J GrönroosJie LiuLin WanXiao LiangYiqing LingYuexuan HuaAnne RoivainenJuhani KnuutiQingguo XieMika TeräsNicola D'AscenzoRiku Klén
Published in: EJNMMI research (2020)
The systems performed most optimal in terms of NEMA image quality parameters when using 18F, where 11C and 89Zr performed slightly worse than 18F. The performance was least optimal when using 68Ga, due to large positron range. The large quantification differences prompt optimization not only by terms of image quality but also quantification. Further investigation should be performed to find an appropriate calibration and harmonization protocol and the evaluation should be conducted on a multi-scanner and multi-center level.