Accuracy, repeatability, and interplatform reproducibility of T1 quantification methods used for DCE-MRI: Results from a multicenter phantom study.

Octavia BaneStefanie J HectorsMathilde WagnerLori L ArlinghausMadhava P AryalYue CaoThomas L ChenevertFiona FennessyWei HuangNola M HyltonJayashree Kalpathy-CramerKathryn E KeenanDariya I MalyarenkoRobert V MulkernDavid C NewittStephen E RussekKarl F StupicAlina TudoricaLisa J WilmesThomas E YankeelovYi-Fei YenMichael A BossBachir Taouli
Published in: Magnetic resonance in medicine (2017)
The VFA protocols with 2 to 3 flip angles optimized for different applications achieved acceptable balance of extensive spatial coverage, accuracy, and repeatability in T1 quantification (errors < 15%). Further optimization in terms of flip-angle choice for each tissue application, and the use of B1 correction, are needed to improve the robustness of VFA protocols for T1 mapping. Magn Reson Med 79:2564-2575, 2018. © 2017 International Society for Magnetic Resonance in Medicine.