Multicentre external validation of a commercial artificial intelligence software to analyse chest radiographs in health screening environments with low disease prevalence.

Cherry KimZepa YangSeong Ho ParkSung Ho HwangYu-Whan OhEun-Young KangHwan Seok Yong
Published in: European radiology (2023)
• This study shows the limited applicability of commercial AI software for detecting abnormalities in CXRs in a health screening population. • When using AI software in a specific clinical setting that differs from the training setting, it is necessary to adjust the threshold or perform additional training with such data that reflects this environment well. • Prospective test accuracy studies, randomised controlled trials, or cohort studies are needed to examine AI software to be implemented in real clinical practice.