Preliminary comparisons between a point-of-care ketometer and reference method using Steller sea lion pup whole blood and plasma.

Stephanie G CrawfordRobert H CokerLorrie D Rea
Published in: Conservation physiology (2024)
We evaluated the Precision Xtra™ ketometer as part of a larger study categorizing fasting status of free-ranging Steller sea lion ( Eumetopias jubatus ; SSL) pups which necessitated the identification of plasma β-hydroxybutyrate concentrations ([β-HBA]) around a threshold of <0.3 and ≥0.3 mmol/l. Whole blood samples mixed with sodium heparin (NaHep) or ethylenediaminetetraacetic acid liquid anticoagulants were tested <10 minutes after collection ( n  = 14; triplicate technical replicates). Plasma (stored at -80°C, NaHep, Thaw1 ) measured via our laboratory's Reference Assay (Sigma Aldrich, St. Louis, MO, Kit #MAK041) served as the standard [β-HBA] for ketometer comparisons. Our observed β-HBA range (0.0-1.6 mmol/l), consistent with published [β-HBA] of free-ranging Otariid pups, represented the lower 20% of the ketometer's range (0.0-8.0 mmol/l). The maximal coefficient of variation (%CV) of ketometer technical replicates was 9.1% (NaHep, whole blood). The majority of ketometer technical replicate sets (84%, including all matrices, anticoagulants and thawings) were identical (CV = 0%). We found linear relationships and agreement of ketometer [β-HBA] between whole blood preserved with different anticoagulants and between whole blood and plasma ( Thaw1 ) measurements. The ketometer produced results with linearity to the Reference Assay for both whole blood and plasma ( Thaw1 ). We identified a non-linear relationship between plasma at Thaw1 and Thaw2 (tested four months apart, NaHep), as only samples with higher SSL [β-HBA] decreased in concentration, and all others remained the same. With respect to categorizing SSL pup fasting in our larger study, the ketometer's %Accuracy , % Sensitivity and % Specificity for samples with Reference Assay β-HBA <0.2 and >0.4 mmol/l were 100%. We adopted a modified procedure: plasma samples with mean ketometer concentrations ±0.1 mmol/l of 0.3 mmol/l β-HBA were re-evaluated using the Reference Assay , improving measurement precision from tenths (ketometer) to thousandths (assay) mmol/l. The Precision Xtra™ ketometer was valuable to our application over the range of [β-HBA] observed in SSL pup plasma and whole blood samples.
Keyphrases