Login / Signup

Investigation of Diffusion Characteristics through Microfluidic Channels for Passive Drug Delivery Applications.

Marcus J GoudieAlyssa P GhumanStephanie B CollinsRamana M PidapartiHitesh Handa
Published in: Journal of drug delivery (2016)
Microfluidics has many drug delivery applications due to the ability to easily create complex device designs with feature sizes reaching down to the 10s of microns. In this work, three different microchannel designs for an implantable device are investigated for treatment of ocular diseases such as glaucoma, age-related macular degeneration (AMD), and diabetic retinopathy. Devices were fabricated using polydimethylsiloxane (PDMS) and soft lithography techniques, where surface chemistry of the channels was altered using 2-[methoxy(polyethyleneoxy)propyl]trimethoxysilane (PEG-silane). An estimated delivery rate for a number of common drugs was approximated for each device through the ratio of the diffusion coefficients for the dye and the respective drug. The delivery rate of the model drugs was maintained at a physiological condition and the effects of channel design and surface chemistry on the delivery rate of the model drugs were recorded over a two-week period. Results showed that the surface chemistry of the device had no significant effect on the delivery rate of the model drugs. All designs were successful in delivering a constant daily dose for each model drug.
Keyphrases