Evaluation of intraoperative neuromonitoring (IONM) data with the Mainz IONM Quality Assurance and Analysis tool.
Thomas J MusholtJulia I Staubitz-VernazzaPetra B MusholtPublished in: BJS open (2023)
Due to high error frequencies, IONM data should undergo in-depth review and multi-step cleaning processes before analysis to standardize scientific reporting. Device software calculates latencies differently; therefore reference values are device-specific (latency) and/or set-up-specific (amplitude). Novel C2-specific reference values for latency and amplitude deviate considerably from published values.