Relative profile measurements in 1.5T MR-linacs: investigation of central axis deviations.

Anastasios EpiskopakisVasiliki MargaroniPantelis KaraiskosEfi KoutsouveliNikolas MarinosEleftherios P Pappas
Published in: Physics in medicine and biology (2024)
Objective . In 1.5 T MR-linacs, the absorbed dose central axis (CAX) deviates from the beam's CAX due to inherent profile asymmetry. In addition, a measured CAX deviation may be biased due to potential lateral (to the beam) effective point of measurement (EPOML) shifts of the detector employed. By investigating CAX deviations, the scope of this study is to determine a set ofEPOMLshifts for profile measurements in 1.5 T MR-linacs. Approach . The Semiflex 3D ion chamber and microDiamond detector (PTW, Germany) were considered in the experimental study while three more detectors were included in the Monte Carlo (MC) study. CAX deviations in the crossline and inline profiles were calculated based on inflection points of the 10×10 cm 2 field, at five centers. In MC simulations, the experimental setup was reproduced. A small water voxel was simulated to calculate CAX deviation without the impact of the detector-specificEPOMLshift. Main results . All measurements were consistent among the five centers. MC-based and experimental measurements were in agreement within uncertainties. Placing the microDiamond in the vertical orientation does not appear to affect the detector'sEPOML, which is on its central longitudinal axis. For the Semiflex 3D in the crossline direction, the CAX deviation was 2.3 mm, i.e. 1 mm larger than the ones measured using the microDiamond and simulated considering the ideal water detector. Thus, anEPOMLshift of 1 mm is recommended for crossline profile measurements under both Semiflex 3D orientations. For the inline profile, anEPOMLshift of -0.5 mm was determined only for the parallel configuration. In the MC study, CAX deviations were found detector- and orientation-dependent. The dead volume is responsible for theEPOMLshift only in the inline profile and under the parallel orientation. Significance . This work contributes to data availability on the correction or mitigation of the magnetic field-induced changes in the detectors' response.
Keyphrases
  • monte carlo
  • contrast enhanced
  • magnetic resonance imaging
  • cross sectional
  • electronic health record
  • deep learning