Login / Signup

Evaluation of correlation between Diopsys® NOVA™ fixed-luminance flicker ERG and Diagnosys® Espion 2™ flicker ERG parameters.

Jonathan RegenoldHien Luong DoanHashem GhorabaHassan KhojastehJaclyn Joyce Jaclyn HwangNegin YavariAmir AkhavanrezayatNgoc Trong Tuong ThanAnthony Huy Dinh LeMuhammad Sohail HalimQuan Dong Nguyen
Published in: Documenta ophthalmologica. Advances in ophthalmology (2023)
There is a statistically significant positive correlation between light-adapted Diopsys® NOVA™ fixed-luminance flicker amplitude and Diagnosys® flicker magnitude values. Additionally, there is a statistically significant positive correlation between Diopsys® NOVA™ fixed-luminance flicker implicit time (converted from phase) and Diagnosys® flicker implicit time values. These results imply that the Diopsys® NOVA™ module, which utilizes the nonstandard shortened International Society for Clinical Electrophysiology of Vision (ISCEV) ERG protocol, can produce reliable light-adapted flicker ffERG measurements.
Keyphrases
  • randomized controlled trial