Point-of-care measurement of fetal blood lactate - Time to trust a new device.
Mandy WangSeng C ChuaLilain BouhadirErin L TreadwellEmma GibbsTherese M McGeePublished in: The Australian & New Zealand journal of obstetrics & gynaecology (2017)
StatStrip appears superior to Lactate Pro 2 to replace the original Lactate Pro. Using StatStrip, the 4.8 mmol/L intervention threshold equivalent was 4.9-5.0 mmol/L. The variation in accuracy of point-of-care lactate devices may exceed the small increments (eg <4.2 mmol/L vs >4.8 mmol/L) that guide obstetric decisions.