Compositional and beam-size-dependent effects on pressure baseline in clumped isotope mass spectrometry.

Ryan A VenturelliBrad E Rosenheim
Published in: Rapid communications in mass spectrometry : RCM (2018)
A seemingly small change in our isotope ratio mass spectrometer resulted in a better understanding of PBL, for which we have developed an empirically based correction protocol to apply. Our new protocol has the potential to reduce analytical time for laboratories measuring PBL, and supports the need for carbonate mineral-based clumped isotope standards.