Evaluation of hernia surgical meshes sterilised with ethylene oxide for adoption under UK regulations.

Alessandra GrilloZargham HyderVivek MuderaAlvena Kureshi
Published in: Surgical endoscopy (2023)
We demonstrated that EO sterilisation affects LCM mechanical properties, but they still have values closer to the native tissues than the commercially available ones. We also showed that in vitro biocompatibility of LCM2 is not affected by EO sterilisation, as HDFs attached and proliferated on the mesh, while EO affected attachment on LCM1. A more detailed cost analysis of the potential savings for healthcare systems around the world needs to be performed to strengthen the cost-effectiveness of this frugal innovation.