Delayed pericardial effusion after left atrial appendage closure with the LAmbre device: Importance of a fully open umbrella.
Fang-Yi XiaoYanyan ChenYi-He ChenXiaodong ZhouXinlei WuXiao ChenLiangguo WangYing FangLan SuWei-Jian HuangPublished in: Journal of cardiovascular electrophysiology (2021)
LAAC with the LAmbre device using an MM significantly increases the rate of a fully open umbrella and decreases the requirement for recapture and the incidence of delayed PE/PT. This method is more effective in patients with combined treatment.