Login / Signup

In vivo human head MRI at 10.5T: A radiofrequency safety study and preliminary imaging results.

Alireza Sadeghi-TarakamehLance DelaBarreRussell L LagoreAngel Torrado-CarvajalXiaoping WuAndrea GrantGregor AdrianyGregory J MetzgerPierre-François Van de MoorteleKamil UgurbilErgin AtalarYigitcan Eryaman
Published in: Magnetic resonance in medicine (2019)
In this study, we acquired the first in vivo human head images at 10.5T using an 8-channel transmit/receive coil. We implemented and expanded a previously proposed workflow to validate the electromagnetic simulation model of the 8-channel transmit/receive coil. Using the validated coil model, we calculated radiofrequency power levels to safely image human subjects.
Keyphrases
  • endothelial cells
  • induced pluripotent stem cells
  • pluripotent stem cells
  • deep learning
  • magnetic resonance imaging
  • high resolution
  • computed tomography
  • high frequency
  • convolutional neural network