Proton-decoupled carbon magnetic resonance spectroscopy in human calf muscles at 7 T using a multi-channel radiofrequency coil.

Sigrun GoluchRoberta Frass-KrieglMartin MeyerspeerMichael PichlerJürgen SiegMartin GajdošíkMartin KrššákElmar Laistler
Published in: Scientific reports (2018)
13C magnetic resonance spectroscopy is a viable, non-invasive method to study cell metabolism in skeletal muscles. However, MR sensitivity of 13C is inherently low, which can be overcome by applying a higher static magnetic field strength together with radiofrequency coil arrays instead of single loop coils or large volume coils, and 1H decoupling, which leads to a simplified spectral pattern. 1H-decoupled 13C-MRS requires RF coils which support both, 1H and 13C, Larmor frequencies with sufficient electromagnetic isolation between the pathways of the two frequencies. We present the development, evaluation, and first in vivo measurement with a 7 T 3-channel 13C and 4-channel 1H transceiver array optimized for 1H-decoupled 13C-MRS in the posterior human calf muscles. To ensure minimal cross-coupling between 13C and 1H arrays, several strategies were combined: mutual magnetic flux was minimized by coil geometry, two LCC traps were inserted into each 13C element, and band-pass and low-pass filters were integrated along the signal pathways. The developed coil array was successfully tested in phantom and in vivo MR experiments, showing a simplified spectral pattern and increase in signal-to-noise ratio of approximately a factor 2 between non-decoupled and 1H-decoupled spectra in a glucose phantom and the human calf muscle.