Login / Signup

Comparative analysis of in vitro accelerated glistening formation in foldable hydrophobic intraocular lenses.

Tamer TandoganGerd Uwe AuffarthChul Young ChoiHyeck-Soo SonRamin Khoramnia
Published in: International ophthalmology (2021)
Glistenings could be induced in all studied IOLs using the accelerated laboratory method. The Alcon AcrySof ® MA60AC and HOYA iSert® PC-60AD IOLs showed MV of high density, while the glistenings in the Hoya IOLs were smaller in size compared to the Alcon IOLs. The MV density was minimal in the Bausch&Lomb enVista and Kowa Avansee™ PU6A IOLs. The propensity of the Alcon AcrySof ® MA60AC IOLs to form glistenings in vitro correlated with the findings of clinical results that are already published.
Keyphrases
  • high density
  • randomized controlled trial
  • high glucose
  • diabetic rats
  • cataract surgery
  • oxidative stress
  • meta analyses