No difference in component migration at five years between the cemented cruciate-retaining ATTUNE and PFC-Sigma knee prosthesis: an update of a randomized clinical radiostereometry trial.

Yasin YükselLennard A KosterBart L KapteinRob G H H NelissenPeter den Hollander
Published in: The bone & joint journal (2023)
MTPM migration at five-year follow-up of the femoral and tibial component of the ATTUNE were similar and as low as that of the PFC-Sigma. MTPM migration of both knee implants did not significantly change from one year post-surgery, indicating stable fixation. Long-term ATTUNE performance may be expected to be comparable to the clinically well-performing PFC-Sigma. We have not found evidence of increased tibial component migration as measured by RSA to support concerns about cement debonding and a higher risk of aseptic loosening with the ATTUNE TKA.
Keyphrases
  • total knee arthroplasty
  • total hip
  • minimally invasive
  • randomized controlled trial
  • clinical trial
  • anterior cruciate ligament reconstruction
  • coronary artery bypass
  • coronary artery disease
  • phase iii
  • open label