Worldwide pacemaker and defibrillator reuse: Systematic review and meta-analysis of contemporary trials.

Sunil K SinhaBhradeev SivasambuGayane YenokyanThomas C CrawfordJonathan ChrispinKim A EagleAndreas S BarthJohn Jack RickardDavid D SpraggStephen C VlayRonald BergerCharles J LoveHugh CalkinsGordon F TomaselliJoseph E Marine
Published in: Pacing and clinical electrophysiology : PACE (2018)
Device reuse utilizing modern protocols did not significantly increase risk of infection, malfunction, premature battery depletion, or device-related death in observational studies. These data provide rationale for proceeding with a prospective multicenter noninferiority randomized control trial.