Login / Signup

Paper-Based Potentiometric Sensors for Nicotine Determination in Smokers' Sweat.

Abde El-Galil E AmrAyman H KamelAbdulrahman A AlmehiziaAhmed Y A SayedElsayed A ElsayedHisham S M Abd-Rabboh
Published in: ACS omega (2021)
Herein, we describe for the first time, the design and fabrication of a novel nicotine paper-based sensor, in which a miniaturized paper reference electrode is integrated for potentiometric measurements. The paper-based sensors were designed using printed wax barriers to define the electrochemical cell and the sample zones. The electrodes were based on the use of the ion association complexes of the nicotinium cation (Nic) with either tetraphenylborate (TPB) or 5-nitrobarbiturate (NB) counter anions as sensing materials for nicotine recognition. A poly (3,4 ethylenedioxythiophene)/poly-(styrene sulfonate) (PEDOT/PSS) conducting polymer was used as an ion-to-electron transducer. The performance characteristics of the proposed sensors were evaluated and it revealed a rapid and stable response with a Nernstian slope of 55.2 ± 0.3 and 51.2 ± 0.6 mV/decade over the linear range of 1.0 × 10-5 to 1.0 × 10-2 M and detection limits of 6.0 and 8.0 μM for [Nic/TPB] and [Nic/NB], respectively. The sensors revealed a constant response over the pH range 3.5-6.5. The designed sensors provided a portable, inexpensive, and disposable way of measuring trace levels of nicotine coming from different cigarettes and in the collected human sweat of heavy smokers. All results were compared favorably with those obtained by the standard gas chromatographic method.
Keyphrases