Login / Signup

The effect of undersizing and tapping on bone to implant contact and implant primary stability: A histomorphometric study on bovine ribs.

Danilo Alessio Di StefanoVittoria PerrottiGian Battista GrecoClaudia CappucciPaolo ArosioAdriano PiattelliGiovanna Iezzi
Published in: The journal of advanced prosthodontics (2018)
The integral of the torque-depth curve provided reliable information about bone-to-implant contact and primary implant stability even in tapped or undersized sites. The linear relations found among the parameters suggests a connection between extent and modality of undersizing and the corresponding increase of the integral and, consequently, of primary stability. These results might help the physician determine the extent of undersizing needed to achieve the proper implant primary stability, according to the planned loading protocol.
Keyphrases
  • soft tissue
  • bone mineral density
  • randomized controlled trial
  • postmenopausal women