Login / Signup

Semiautomatic cochleostomy target and insertion trajectory planning for minimally invasive cochlear implantation.

Wilhelm WimmerFrédéric VenailTom WilliamsonMohamed AkkariNicolas GerberStefan WeberMarco CaversaccioAlain UzielBrett Bell
Published in: BioMed research international (2014)
A major component of minimally invasive cochlear implantation is atraumatic scala tympani (ST) placement of the electrode array. This work reports on a semiautomatic planning paradigm that uses anatomical landmarks and cochlear surface models for cochleostomy target and insertion trajectory computation. The method was validated in a human whole head cadaver model (n = 10 ears). Cochleostomy targets were generated from an automated script and used for consecutive planning of a direct cochlear access (DCA) drill trajectory from the mastoid surface to the inner ear. An image-guided robotic system was used to perform both, DCA and cochleostomy drilling. Nine of 10 implanted specimens showed complete ST placement. One case of scala vestibuli insertion occurred due to a registration/drilling error of 0.79 mm. The presented approach indicates that a safe cochleostomy target and insertion trajectory can be planned using conventional clinical imaging modalities, which lack sufficient resolution to identify the basilar membrane.
Keyphrases
  • minimally invasive
  • high resolution
  • endothelial cells
  • ultrasound guided
  • high throughput
  • optic nerve
  • mass spectrometry
  • single cell
  • pluripotent stem cells
  • fine needle aspiration
  • solid state