Accuracy of point-of-care testing devices for haemoglobin in the operating room: meta-analysis.

Hilalion San AhnTori LenetRichard W D GilbertRanjeeta MallickJulie L V ShawDean A FergussonDaniel I McIsaacGuillaume Martel
Published in: BJS open (2024)
All POCT examining intraoperative haemoglobin measurement yielded pooled mean difference LOAs larger than the allowable limit difference of ±4 g/dl. Intraoperative haemoglobin measured by POCT should not be considered interchangeable with central laboratory values and caution is necessary when using these tests to guide intraoperative transfusion.