Login / Signup

Coherence and reliability of a wearable inertial measurement unit for measuring postural sway.

Eva Ekvall HanssonÅsa Tornberg
Published in: BMC research notes (2019)
The coherence between the IMU and the force plate was 0.84 (strong) in the medio-lateral direction with eyes open (EO) and 0.88 (strong) with eyes closed (EC). The ICC for intra-trial reliability for the IMU varied between 0.50 and 0.67 (moderate to good) with a CV between 17.8 and 22.1% and ICC varied between 0.75 and 0.86 (good) for inter-trial reliability, with an SEM of 0.98 to 1.96 mm/s. We have demonstrated that the IMU was both reliable and highly coherent with golden standard, although the two assessment methods were not interchangeable. The ability to move the balance lab out into real life in the form of a wearable device will provide opportunities to perform research that has not been possible before.
Keyphrases
  • optical coherence tomography
  • study protocol
  • phase iii
  • minimally invasive
  • clinical trial
  • phase ii
  • heart rate
  • randomized controlled trial
  • single molecule
  • open label
  • cataract surgery