Login / Signup

Preclinical Studies of the Nebulized Delivery of Liposomal Amphotericin B.

Sagar KothariSteven G KefalosNicholas D HagesTimothy E CorcoranShahid Husain
Published in: Journal of aerosol medicine and pulmonary drug delivery (2022)
Background: Intravenous liposomal amphotericin B (L-AMB) has accompanying side effects that may be diminished when administering an inhaled form. Delivery systems for inhaled or aerosolized L-AMB vary, and there has not been a recent comparison of available systems to date. Methods: We compared three differently designed nebulizer delivery systems for the inhaled delivery of L-AMB to determine the best combination of efficient lung dosing and treatment time. Aerosol size was measured using a Malvern Mastersizer, and five separate nebulizers were tested. For drug output measurements, a Harvard Lung was used, and aerosol was collected using HEPA filters. Results: Overall aerosol size characteristics were similar for all devices with volume median diameters in the 4-5 μm range. The highest inhaled dose was delivered by the AeroEclipse. The Aerogen and the AeroEclipse had similar predicted pulmonary doses, and the AeroEclipse had the highest pulmonary delivery rates. Conclusion: The AeroEclipse nebulizer may provide more efficient delivery in a shorter amount of time; however, human studies are warranted to assess the safety, tolerability, and efficacy of inhaled delivery of L-AMB from this system.
Keyphrases
  • cystic fibrosis
  • pulmonary hypertension
  • endothelial cells
  • randomized controlled trial
  • clinical trial
  • low dose
  • open label
  • mesenchymal stem cells
  • case control
  • double blind
  • drug induced
  • study protocol