Determination of Clinically Acceptable Analytical Variation of Cardiac Troponin at Decision Thresholds.

John William PickeringPeter A KavsakRobert H ChristensonRichard W TroughtonChristopher J PembertonA Mark RichardsLaura R JoyceMartin P Than
Published in: Clinical chemistry (2024)
At the low-risk threshold, analytical variation up to 3 ng/L minimally impacted the safety metric (sensitivity) but marginally reduced effectiveness. Similarly, at the URL even relative variation up to 25% minimally impacted safety metrics and effectiveness. Analytical variation for delta thresholds did not negatively impact sensitivity but decreased effectiveness.
Keyphrases
  • randomized controlled trial
  • systematic review
  • liquid chromatography
  • decision making
  • molecularly imprinted