Login / Signup

Root cause analysis of epidural spinal cord stimulator implant infections with resolution after implementation of an improved protocol for surgical placement.

Forest W ArnoldSarah BishopDavid JohnsonLaShawn ScottCrystal HeishmanLeah OppyTyler BallMayur SharmaClaudia AngeliChristie FerreiraYangsheng ChenSusan HarkemaMaxwell Boakye
Published in: Journal of infection prevention (2019)
The root cause analysis provides guidance for other institutions performing the same novel procedure. This analysis did not reveal a direct association, but did generate several areas for improvement including increasing pre-surgical screening, cleaning transient equipment (e.g., computer screens), limiting traffic in the operating room, using new sterile instruments for each stage of the procedure, not reopening the back incision, not applying vancomycin powder, and using an antimicrobial envelope for the stimulator.
Keyphrases