Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS).

Farahnaz JoukarSara YeganehMohammadreza NaghipourSoheil HassanipourHossein-Ali NikbakhtFariborz Mansour-Ghanaei
Published in: Medical devices (Auckland, N.Z.) (2020)
Omron HBP-1100-E professional blood pressure monitor is not desirable for measuring the BP for Iranian adults as it overestimates blood pressure in this population.
Keyphrases
  • blood pressure
  • hypertensive patients
  • heart rate
  • randomized controlled trial
  • healthcare
  • blood glucose
  • type diabetes
  • metabolic syndrome
  • adipose tissue