Login / Signup

Development of a Novel Nordic Hamstring Exercise Performance Test Device: A Reliability and Intervention Study.

Jesper AugustssonSofia Ryman Augustsson
Published in: Sports (Basel, Switzerland) (2022)
There is evidence that a knee flexor exercise, the Nordic hamstring exercise (NHE), prevents hamstring injuries. The purpose of this study was therefore to develop, and to determine the reliability of, a novel NHE test device and, further, to determine the effectiveness of a 10-week low volume NHE program on NHE performance. Twenty female football (soccer) players, aged 16-30 years, participated in this study. From a kneeling position on the device, with the ankles secured under a heavy lifting sling, participants leaned forward in a controlled manner as far as possible (eccentric phase) and then returned to the starting position (concentric phase). A tape measure documented the forward distance achieved by the participants in cm. Participants completed three separate occasions to evaluate test-retest reliability. Additionally, 14 players performed a low volume (1 set of 5 repetitions) NHE program once weekly for 10 weeks. No significant test-retest differences in NHE performance were observed. The intra-class correlation coefficient was 0.95 and the coefficient of variation was 3.54% between tests. Mean improvement in the NHE performance test by the players following training was 22% (8.7 cm), p = 0.005. Our test device reliably measured NHE performance and is easy to perform in any setting. Further, NHE performance was improved by a 10-week low volume NHE program. This suggests that even a small dose (1 set of 5 repetitions once weekly) of the NHE may enhance NHE performance.
Keyphrases