Phantom assessment of three-dimensional geometric distortion of a dedicated wide-bore MR-simulator for radiotherapy.

Max W L LawJing YuanOi Lei WongAbby D YingYihang ZhouKin Yin CheungSiu Ki Yu
Published in: Biomedical physics & engineering express (2021)
This study evaluated the machine-dependent three-dimensional geometric distortion images acquired from a 1.5T 700mm-wide bore MR-simulator based on a large geometric accuracy phantom. With the consideration of radiation therapy (RT) application requirements, every sequence was examined in various combinations of acquisition-orientations and receiver-bandwidths with console-integrated distortion correction enabled. Distortion was repeatedly measured over a six-month period. The distortion measured from the images acquired at the beginning of this period was employed to retrospectively correct the distortion in the subsequent acquisitions. Geometric distortion was analyzed within the largest field-of-view allowed. Six sequences were examined for comprehensive distortion analysis - VIBE, SPACE, TSE, FLASH, BLADE and PETRA. Based on optimal acquisition parameters, their diameter-sphere-volumes (DSVs) of CT-comparable geometric fidelity (where 1mm distortion was allowed) were 333.6mm, 315.1mm, 316.0mm, 318.9mm, 306.2mm and 314.5mm respectively. This was a significant increase from 254.0mm, 245.5mm, 228.9mm, 256.6mm, 230.8mm and 254.2mm DSVs respectively, when images were acquired using un-optimized parameters. The longitudinal stability of geometric distortion and the efficacy of retrospective correction of console-corrected images, based on prior distortion measurements, were inspected using VIBE and SPACE. The retrospectively corrected images achieved over 500mm DSVs with 1mm distortion allowed. The median distortion was below 1mm after retrospective correction, proving that obtaining prior distortion map for subsequent retrospective distortion correction is beneficial. The systematic evaluation of distortion using various combinations of sequence-type, acquisition-orientation and receiver-bandwidth in a six-month time span would be a valuable guideline for optimizing sequence for various RT applications.
Keyphrases