Assessment of the performance of blood glucose monitoring systems for monitoring dysglycaemia in neonatal patients.

Yin BaJin XuLin YuanHaiyan ZhuYipei YangMei Mei LamShulian ZhangMili XiaoJiayin XuRong ZhangChao Chen
Published in: BMJ paediatrics open (2018)
This three-step validation protocol provides a challenging approach for determining the accuracy and reliability of BGMS for managing dysglycaemia in neonates. StatStrip BGMS achieved analytical and clinical performance criteria confirming its suitability for use in neonates. We advocate that this validation approach should be considered for performance evaluations of both BGMS and continuous glucose monitoring systems going forward.