Login / Signup

Validity of the Polar Vantage M watch when measuring heart rate at different exercise intensities.

Tricia ShumateMagdalen LinkJames W FurnessKevin Kemp-SmithVinicius SimasMike Climstein
Published in: PeerJ (2021)
This study revealed that the PVM watch had a strong correlation with the ECG throughout the entire Bruce Protocol, however the level of agreement (LoA) becomes widely dispersed as exercise intensities increased. Due to the large LoA between the ECG and PVM watch, it is not advisable to use this device in clinical populations in which accurate HR measures are essential for patient safety; however, the watch maybe used in settings where less accurate HR is not critical to an individual's safety while exercising.