Developing a surface acoustic wave-induced microfluidic cell lysis device for point-of-care DNA amplification.
Abbas Ali HusseiniAli Mohammad YazdaniFatemeh GhadiriAlper ŞişmanPublished in: Engineering in life sciences (2023)
We developed a microchip device using surface acoustic waves (SAW) and sharp-edge glass microparticles to rapidly lyse low-level cell samples. This microchip features a 13-finger pair interdigital transducer (IDT) with a 30-degree focused angle, creating high-intensity acoustic beams converging 6 mm away at a 16 MHz frequency. Cell lysis is achieved through centrifugal forces acting on Candida albicans cells and glass particles within the focal area. To optimize this SAW-induced streaming, we conducted 42 pilot experiments, varying electrical power, droplet volume, glass particle size, concentration, and lysis time, resulting in optimal conditions: an electrical signal of 2.5 W, a 20 μL sample volume, glass particle size below 10 μm, concentration of 0.2 μg, and a 5-min lysis period. We successfully amplified DNA target fragments directly from the lysate, demonstrating an efficient microchip-based cell lysis method. When combined with an isothermal amplification technique, this technology holds promise for rapid point-of-care (POC) applications.
Keyphrases
- single cell
- high intensity
- candida albicans
- circulating tumor
- randomized controlled trial
- high glucose
- stem cells
- machine learning
- biofilm formation
- high resolution
- cell death
- mass spectrometry
- resistance training
- cell free
- deep learning
- label free
- endoplasmic reticulum stress
- stress induced
- cell cycle arrest
- loop mediated isothermal amplification